| Error Code | Description | Probable Cause | Recommended Action | |------------|-------------|----------------|---------------------| | | C-arm rotation encoder failure | Broken encoder or loose belt | Replace position encoder | | E204 | Brake release timeout | Brake solenoid or relay stuck | Check brake PCB and 24V supply | | E210 | Collimator homing error | Limit switch or stepper motor failure | Manually move collimator; test motor |
This comprehensive guide is designed to help biomedical engineers, radiology technicians, and service personnel understand, diagnose, and resolve common .
Always document the exact error code, frequency, and environmental conditions (mains voltage, room temperature) before calling technical support.
The Siremobil Compact L uses oil cooling for the X-ray tube housing and heat exchanger.