The core technology of the Surgitron Radiolase is its high-frequency, low-temperature waveform. Traditional electrosurgical units (ESUs) operate at lower frequencies (500 kHz to 1 MHz), which often cause significant lateral heat damage to surrounding tissue. ellman surgitron radiolase service manual
| Symptom | Probable cause | |---------|----------------| | No power at all | Main fuse (rear), line filter, power switch | | Front panel lights on, no RF | Footswitch relay stuck/open; output transistor (IRF510 or similar) | | Cut works, coag weak | Bad ceramic cap in coag mode switching circuit | | Continuous RF on power-up | Shorted MOSFET in output stage | Instructions for vs
A genuine service manual (Ellman part number 230-0115-1 for older models or 230-0159 for the Radiolase II) is typically divided into five major sections. no RF | Footswitch relay stuck/open